• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TPRLC XR FP TYPE1 PPS 7X134MM 1; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TPRLC XR FP TYPE1 PPS 7X134MM 1; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Tear, Rip or Hole in Device Packaging (2385)
Patient Problem No Patient Involvement (2645)
Event Date 03/23/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).(b)(6).Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2020-01770, 0001825034-2020-01771, 0001825034-2020-01772.Reported event was confirmed by review of photographs.Device history record (dhr) was reviewed and no discrepancies were found.The root cause of the reported issue is attributed to transit damage.Customer has indicated that the product will be returned to the manufacturer for evaluation, however, the product has yet to be received.Alternately, evaluation of the photographs provided confirmed that the sterile packaging (blister) is damaged.The likely condition of the parts when they left zimmer biomet control is considered conforming based on the device history record (dhr) review.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that during investigation of circulated items, devices were identified with damage to their sterile packages.No patients were involved.No additional information available.
 
Event Description
Not reportable: upon investigation, it has been determined that the damage to the packaging did not breach the sterility.Event is no longer considered reportable for this device, and initial report should be voided.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.Reported event was confirmed by visual examination of the device.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was traced to transport/storage issues.Products have been returned.Visual evaluation of the returned products identified that the outer sterile cavities have been damaged.Sterility has not been compromised.Not reportable: upon investigation, it has been determined that the damage to the pacakaging did not breach the sterility.Event is no longer considered reportable for this device, and initial report should be voided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC XR FP TYPE1 PPS 7X134MM 1
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key10013280
MDR Text Key189579246
Report Number0001825034-2020-01773
Device Sequence Number1
Product Code LPH
UDI-Device Identifier00880304516908
UDI-Public00880304516908
Combination Product (y/n)N
PMA/PMN Number
K120030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 06/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number51-102070
Device Lot Number3735251
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/13/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/23/2020
Initial Date FDA Received04/30/2020
Supplement Dates Manufacturer Received06/04/2020
Supplement Dates FDA Received06/10/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-