Qn#(b)(4).A device history record review was performed on the kit and ampules with a potentially relevant finding.For material # k-07900-001 (lidocaine w/epinephrine 5 ml), lot # 23p19l0082, according to incoming inspection records, 1 of 315 ampules were observed broken in a batch of 19200.This is outside of the parameter for this defect.A corrective action is not required at this time as a potential root cause could not be determined based upon the information provided and without a sample.A corrective action is not required at this time as a potential root cause could not be determined based upon the information provided and without a sample.
|