Catalog Number H938738
Device Problem Material Split, Cut or Torn (4008)
Patient Problem No Patient Involvement (2645)
Event Date 04/08/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a 500ml eva (ethyl vinyl acetate) tpn (total parenteral nutrition) bag was damaged.The damage was further described as "rupture next to the green closure".This was identified prior to use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information was added to h3, h4 and h6.H4: the lot was manufactured from april 22, 2019 - april 23, 2019.H10: the actual device was not available; however, a companion sample was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.A functional testing was performed using tap water and no issues were noted.The reported condition was not verified on the companion sample.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
500 ML TPN BAG
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key10017674
MDR Text Key189701456
Report Number1416980-2020-02483
Device Sequence Number1
Product Code LHI
UDI-Device Identifier00085412477220
UDI-Public(01)00085412477220
Combination Product (y/n)N
PMA/PMN Number
K900585
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 06/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date03/31/2022
Device Catalogue NumberH938738
Device Lot Number60181158
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/08/2020
Initial Date FDA Received05/01/2020
Supplement Dates Manufacturer Received06/06/2020
Supplement Dates FDA Received06/08/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-