Model Number 407200
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/07/2020
Event Type  malfunction  
Manufacturer Narrative
One brk transseptal needle was received for evaluation.The needle tip had been fractured, detached, and was not returned.Functional testing was not possible due to the aforementioned damage.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the received condition of the device and the information provided, the cause of the reported event remains unknown.
 
Event Description
This event is being reported based on analysis which revealed the tip was fractured and detached.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRK TRANSSEPTAL NEEDLE, 71 CM LENGTH
Type of Device
TROCAR
Manufacturer (Section D)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
stephanie o' sullivan
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key10038142
MDR Text Key190316371
Report Number3008452825-2020-00217
Device Sequence Number1
Product Code DRC
UDI-Device Identifier05414734205092
UDI-Public05414734205092
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K072278
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 05/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date04/30/2022
Device Model Number407200
Device Catalogue Number407200
Device Lot Number7004520
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/31/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/27/2020
Initial Date FDA Received05/08/2020
Date Device Manufactured05/17/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-