Model Number 3100 A |
Device Problem
Pressure Problem (3012)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 04/14/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Any additional information provided by the customer will be included in a follow up report.At this time, vyaire has not received the suspect device/component for evaluation.Therefore, no root cause can be determined yet.
|
|
Event Description
|
The customer reported to vyaire medical that the 3100a ventilator cannot get to pressurize and the driver would not start.The customer confirmed that there was no patient involvement associated with the reported event.
|
|
Manufacturer Narrative
|
A vyaire medical field service representative (fsr) evaluated the suspect device onsite.The fsr replaced driver and 6k kit.Also replaced driver power module for 7yr preventative maintenance.Calibrated airway pressure transducer, pneumatics, and driver displacement indicator.Checked alarms and ran performance test.Unit passed all test and runs according to manufacture specifications.
|
|
Search Alerts/Recalls
|