• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THORATEC SWITZERLAND GMBH CENTRIMAG MOTOR, US; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THORATEC SWITZERLAND GMBH CENTRIMAG MOTOR, US; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE Back to Search Results
Model Number 102956
Device Problems Pressure Problem (3012); Insufficient Information (3190)
Patient Problems No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/21/2020
Event Type  malfunction  
Manufacturer Narrative
No further information was provided.A supplemental report will be submitted when the manufacturer's investigation is completed.
 
Event Description
Related manufacturer report number: 2916596-2020-02474.It was reported that the centrimag console displayed a pressure system fault alert.
 
Manufacturer Narrative
Manufacturer's investigation conclusion: the reported event of pressure alarms was not confirmed.The centrimag motor was not returned for analysis.A good faith effort was sent to retrieve the serial number of the centrimag motor; however, this information was not received.Additional provided information communicated on 14jul2020 stated there were no patient complications and that the motor was in use and was returned to abbott; however, to date, the motor has yet to be returned to abbott.A good faith effort was sent to request additional information on if the motor will be returned and if there is tracking information available; however, no response has been received.The root cause for the reported event was unable to be conclusively determined through this analysis.The 2nd generation centrimag system operating manual section 4 entitled "warnings & precautions" warns "one additional 2nd generation centrimag primary console, motor and flow probe are required as backup system in the immediate vicinity of each patient whenever the centrimag or pedivas blood pump is used.The backup console must be connected to the backup motor and to the backup flow probe, have a battery charge sufficient for at least one hour of operation, be connected to ac power (except during transport) and be immediately available should the main console, motor or flow probe experience a malfunction." the 2nd generation centrimag system operating manual section 10 entitled "emergency and troubleshooting" states that "the recommended practice whenever there is a 2nd generation centrimag primary console or motor malfunction is to replace the console and motor as a set.Remove the blood pump from the malfunctioning motor and console and place the blood pump in the backup motor and console to continue patient support.Do not exchange individual motors or individual consoles during patient support." the 2nd generation centrimag system operating manual section 12.1 entitled "appendix i ¿ primary console alarms and alerts" contains a list of console alarms and alerts, as well as appropriate operator response to these events.No further information was provided.The manufacturer is closing the file on this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIMAG MOTOR, US
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE
Manufacturer (Section D)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ  CH-8005
MDR Report Key10045037
MDR Text Key191031219
Report Number2916596-2020-02548
Device Sequence Number1
Product Code KFM
Combination Product (y/n)N
PMA/PMN Number
K020271
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 02/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number102956
Device Catalogue Number102956
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/21/2020
Initial Date FDA Received05/11/2020
Supplement Dates Manufacturer Received02/05/2021
Supplement Dates FDA Received02/08/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
CENTRIMAG 2ND GENERATION PRIMARY CONSOLE
-
-