Model Number P7900B100011 |
Device Problem
Self-Activation or Keying (1557)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 05/06/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The hillrom technician found the master control power control board, head actuator and foot actuator needed to be replaced.Per the hillrom service manual, perform annual preventive maintenance procedures to make sure all versacare® bed components are functioning as originally designed.Head section motor: examine the motor for the presence and tightness of the attachment hardware.Replace as necessary.Fully raise and lower the head section.Make sure there is no friction or abnormal noises and no audible overload indication can be heard during the movement.Replace in the event of a malfunction.Make sure all functions on the caregiver control work correctly.Repair or replace the side rails as necessary.Per the hillrom service manual, perform annual preventive maintenance procedures to make sure all versacare® bed components are functioning as originally designed.Hilow motor: examine the motor for the presence and tightness of the attachment hardware.Replace as necessary.Fully raise and lower the bed.If the bed does not fully raise and lower, calibrate the bed position sensor.Make sure there is no friction or abnormal noises and no audible overload indication can be heard during the movement.Make sure the bed not down led lights up on the control pendant and goes out when the bed is in low position.Replace the defective motor(s) in the event of a malfunction.Troubleshoot in the event of a doubt.A search of the hillrom maintenance records did not show hillrom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician replaced the master control power control board, head actuator and foot actuator to resolve the issue.Based on this information, no further action is required.
|
|
Event Description
|
Hillrom received a report from a hillrom technician stating the bed's head high/low and frame high/low were engaging without pressing any buttons on the bed's control panel (self run).The bed was located in the bed shop at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint #(b)(4).
|
|
Manufacturer Narrative
|
The hillrom technician found the master control power control board, head actuator and foot actuator needed to be replaced.Per the hillrom service manual, it is necessary for the centrella¿ bed to have an effective maintenance program.We recommend that you do annual preventive maintenance.Plug the bed into a power outlet.Make sure all functions on the caregiver control panel operate correctly.Repair or replace the siderail if necessary.A search of the hillrom maintenance records did not show hillrom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician replaced the master control power control board, head actuator and foot actuator to resolve the issue.Based on this information, no further action is required.
|
|
Event Description
|
Hillrom received a report from a hillrom technician stating the bed's head high/low and frame high/low were engaging without pressing any buttons on the bed's control panel (self run).The bed was located in the bed shop at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint # (b)(4).
|
|