Model Number GIF-Q260J
Device Problems Partial Blockage (1065); Complete Blockage (1094)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The subject device was not returned to olympus medical systems corp.(omsc) for evaluation.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.The exact cause of the reported event could not be conclusively determined.
 
Event Description
Olympus medical systems corp.(omsc) was informed that the instrument channel and the air/water nozzle of the subject device were clogged with non-olympus drug for embolization, histoacryl, that was used in an endoscopic injection sclerotherapy procedure.There were residue in the instrument channel and the air/water nozzle was clogged completely, in spite of the user facility brushed them.The device had been reprocessed with an olympus automated endoscope reprocessor model oer-3 (not available in the usa).There was no report of patient injury associated with the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS LUCERA GASTROINTESTINAL VIDEOSCOPE
Type of Device
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key10046163
MDR Text Key224679440
Report Number8010047-2020-02642
Device Sequence Number1
Product Code FDS
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K954451
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other
Type of Report Initial
Report Date 05/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGIF-Q260J
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/22/2020
Initial Date FDA Received05/12/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/08/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-