• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number VICMO13.2
Device Problem Contamination /Decontamination Problem (2895)
Patient Problem No Code Available (3191)
Event Date 04/22/2020
Event Type  Injury  
Manufacturer Narrative
This product is manufactured in the u.S.But not marketed in the u.S.(b)(4).No similar complaint was reported for units within the same lot.(b)(4).
 
Event Description
The reporter indicated the surgeon implanted a 13.2mm vicmo13.2 implantable collamer lens, -07.50 diopter, in the patient's eye on (b)(6) 2020.The surgeon noted foreign bodies, similar to icl material on the front surface of the lens, after the lens was injected into the eye.There was no damage to the lens and patient's vision, intraocular pressure and vault were normal.There was some iris depigmentation that had occurred.The lens remained implanted.Additional information has been requested but none has been forthcoming.If additional information is received, a supplemental medwatch will be submitted.
 
Manufacturer Narrative
Device history record (dhr) review: based on the results of the investigation all released devices from the associated work order(s) including the suspected device; have been manufactured within established process parameters; and there is no indication that the manufacturing and processing of the device contributed to the complaint issue.Claim # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Type of Device
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia, ca
MDR Report Key10049232
MDR Text Key190771189
Report Number2023826-2020-01065
Device Sequence Number1
Product Code MTA
Combination Product (y/n)N
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Type of Report Initial,Followup
Report Date 04/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date07/31/2022
Device Model NumberVICMO13.2
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/23/2020
Initial Date FDA Received05/12/2020
Supplement Dates Manufacturer Received06/10/2020
Supplement Dates FDA Received07/08/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-