• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. CARELINK PROGRAMMER; PROGRAMMER, PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. CARELINK PROGRAMMER; PROGRAMMER, PACEMAKER Back to Search Results
Model Number 2090
Device Problems No Display/Image (1183); Mechanical Problem (1384); Overheating of Device (1437); Operating System Becomes Nonfunctional (2996); Power Problem (3010)
Patient Problem No Patient Involvement (2645)
Event Date 05/08/2020
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that upon plugging in and turning on the programmer, the device suddenly stopped working and smoke was emitted from the side fan vent.The programmer was then unplugged.The device is expected to be returned for service.There was no patient involvement.
 
Manufacturer Narrative
Product analysis: analysis was able to confirm the customers comment regarding the programmer.It was found that the device would not power up and a burn odor was noted.The power supply was replaced.Salvaged part was used for power supply.All salvaged part used was inspected per applicable requirements.The hard drive was reconfigured, reloaded, and it's software updated.The programmer passed final functional and system tests.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was further reported that the programmers screen went black.The device has been returned for servicing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK PROGRAMMER
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key10049598
MDR Text Key191550708
Report Number2182208-2020-00895
Device Sequence Number1
Product Code KRG
UDI-Device Identifier00613994581570
UDI-Public00613994581570
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 07/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number2090
Device Catalogue Number2090
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/12/2020
Initial Date Manufacturer Received 05/08/2020
Initial Date FDA Received05/12/2020
Supplement Dates Manufacturer Received06/02/2020
Supplement Dates FDA Received07/01/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/30/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-