MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. IV3000 WINDOW FRAME 6X7CM CTN 100; TAPE AND BANDAGE, ADHESIVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. IV3000 WINDOW FRAME 6X7CM CTN 100; TAPE AND BANDAGE, ADHESIVE Back to Search Results
Catalog Number 59410082
Device Problem Manufacturing, Packaging or Shipping Problem (2975)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/17/2020
Event Type  malfunction  
Event Description
It was reported that the dressing of the iv-3000 was creased, the surrounding protect paper fall off.So the dressing couldn't be used.
 
Manufacturer Narrative
We have now concluded our investigation into this complaint.The batch records were reviewed and it could be confirmed that the products had progressed through a satisfactory release process which involved testing the product against the requirements of its finished product specification.There was no issues recorded in the batch log of the manufacturing records during the manufacturing process that could have caused or contributed to the issue.A complaint history review was carried out using the lot and part numbers provided, there have been no further complaints reported with this failure mode in the past three years.The device was intended for use in treatment.The returned samples were functionally evaluated which confirmed the failure mode of delamination.No creasing was observed.Delamination is a low level intermittent fault which if seen is tabbed for removal during the conversion process.During the manufacturing of the dressing, in process checks are undertaken for this product.All material is manufactured according to the specification and only product meeting the release criteria will be passed on for further processing.Unfortunately on this rare occasion in process checks failed to identify the failure mode and therefore the root cause was a manufacturing process issue.We were able to confirm a relationship between the event and the device.As this is a very low occurring issue, no further actions by smith and nephew are deemed necessary at this stage.However, we will continue to monitor for any adverse trends relating to this product range.Smith and nephew acknowledge customer concern and are continually investigating ways to develop and improve our products.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IV3000 WINDOW FRAME 6X7CM CTN 100
Type of Device
TAPE AND BANDAGE, ADHESIVE
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
MDR Report Key10062523
MDR Text Key191156526
Report Number8043484-2020-00454
Device Sequence Number1
Product Code KGX
UDI-Device Identifier04582111151568
UDI-Public4582111151568
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Type of Report Initial,Followup
Report Date 07/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number59410082
Device Lot Number1809
Was Device Available for Evaluation? Yes
Distributor Facility Aware Date04/20/2020
Initial Date Manufacturer Received 04/20/2020
Initial Date FDA Received05/15/2020
Supplement Dates Manufacturer Received07/16/2020
Supplement Dates FDA Received07/08/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-