• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA III XENON LIGHT SOURCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA III XENON LIGHT SOURCE Back to Search Results
Model Number CLV-190
Device Problem Output below Specifications (3004)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
The asset unit was returned to the service center.A physical evaluation was performed and the locking mechanism of the scope connector socket was worn out.There was damage to the housing of the unit as the top cover had multiple scratches, the rear panel had some rusting.Additionally, a third party lamp was installed and the lamp life read 164+ hours.To mitigate device damage, the instruction manual provides warning which states, "never install a lamp that has not been approved by olympus.The use of a non-approved lamp can cause damage to the light source and ancillary equipment, malfunction or a fire".
 
Event Description
The service center became aware that the xenon light source asset unit was found to have a third party lamp installed.There was no patient involvement reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key10071707
MDR Text Key207820560
Report Number8010047-2020-02786
Device Sequence Number1
Product Code NWB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
CLASS2-EXMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 05/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberCLV-190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/17/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/22/2020
Initial Date FDA Received05/19/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-