BOSTON SCIENTIFIC CORPORATION COMET; TRANSDUCER, PRESSURE, CATHETER TIP Back to Search Results
Model Number 8900
Device Problem Failure to Zero (1683)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/19/2020
Event Type  malfunction  
Manufacturer Narrative
Device evaluated by mfr: the tip, device shaft, sensor port and the coefficient values were examined for damage or any irregularities.The wire shaft showed 1 kink located 162.5cm from the tip and at the occ handle.The tip showed bend damage.There was some peeled coating at the 168.5cm location.The occ cable was connected to the ffr link for signal verification.The signal was not present as designed.Device analysis was conducted by inspecting the proximal end of the wire for any damage to the fiber optic.No damage was noticed.The sensor was inspected by viewing the sensor port to verify that the sensor was in the correct location.This sensor looked to be too far distal which would give the indication of the sensor being detached from the fiber optic cable.The wire was gently shaken to see if the sensor would move within the sensor housing and the sensor did move which verifies the sensor was detached from the fiber optic.The coefficient values were confirmed to be programmed.The wire was removed from the occ handle with no issues.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.Device analysis determined the condition of the returned device was consistent with the reported information of signal/pressure issues.
 
Event Description
Reportable based on analysis completed on 23apr2020.It was reported that the comet wire was not able to zero.This occurred outside the patient and a different device was used to complete the procedure without issue or patient injury.However, returned device analysis revealed peeled coating.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMET
Type of Device
TRANSDUCER, PRESSURE, CATHETER TIP
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key10072475
MDR Text Key191583861
Report Number2134265-2020-06139
Device Sequence Number1
Product Code DXO
UDI-Device Identifier08714729904403
UDI-Public08714729904403
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K151610
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 05/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/09/2022
Device Model Number8900
Device Catalogue Number8900
Device Lot Number0025019088
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/31/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/23/2020
Initial Date FDA Received05/19/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/10/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-