• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOFIRE DIAGNOSTICS, LLC FILMARRAY® BLOOD CULTURE IDENTIFICATION PANEL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOFIRE DIAGNOSTICS, LLC FILMARRAY® BLOOD CULTURE IDENTIFICATION PANEL Back to Search Results
Model Number RFIT-ASY-0126
Device Problem False Positive Result (1227)
Patient Problem Renal Failure (2041)
Event Date 03/15/2020
Event Type  Injury  
Manufacturer Narrative
Investigation: (b)(6) hospital and medical center ((b)(6)) reported potential (b)(6) results on the filmarray® bcid panel after testing a sample from a (b)(6)-year-old, female patient.The patient had symptoms of leukocytosis, hypotension, weakness, fatigue, and acute kidney injury (aki).The sample was cultured using biomerieux bact/alert® sa culture bottles (lot# unknown).The bottle was collected on (b)(6) 2020 at 8:50 pm and rang positive on (b)(6) 2020 at 12:13 pm on (b)(6) 2020, the sample was tested on the filmarray bcid panel and was (b)(6) for staphylococcus aureus and (b)(6).The sample was gram stained and had gram-positive cocci in clusters.The sample was cultured and detected staphylococcus by latex testing.The sample was tested using vitek® ast-gp75 susceptibility card with a cefoxitin screen negative.It was also tested by kb disk diffusion and had a cefoxitin zone of 25mm and was sensitive.The sample was retested on the filmarray bcid panel on (b)(6) 2020 and had a (b)(6) s.Aureus and (b)(6) result.The same sample was used for all testing.Due to the filmarray bcid panel results, the patient was treated with broad coverage vancomycin.The patient was treated for 6 weeks because it was unclear if the patient could safely be de-escalated to a narrower spectrum agent due to the discordant results.The patient experienced a possible adverse treatment affect where the serum creatinine (scr) rose from 1.2 to 2.4 while finishing the vancomycin treatment at an outside facility.According to the customer, the patient's "kidney function was impaired and the vancomycin dose was adjusted accordingly." the customer reported that they did not know if the kidney impairment has been fully resolved as the patient was discharged from the hospital and is being managed by an outside external care facility.The customer stated that they would have treated with a less nephrotoxic drug if the results had been (b)(6) for (b)(6).When asked about their test preparation methods, the customer stated that pouches are setup according to the filmarray bcid panel quick guide.The customer decontaminates the biosafety cabinet work area and pouch-loading stations with 10% bleach followed by a 70% alcohol wipe.This biosafety cabinet is also used for the preparation of nasopharyngeal samples for flu pcr rsv antigen and covid testing as well as stool samples for c.Difficile and norovirus pcr.Quality control (qc) records for pouch lot# 104119 (kit lot#1190119) were reviewed.This pouch lot passed qc criteria and was found within specification.No run malfunction was detected and the fimarray instrument (serial# (b)(4)) was working within design specifications.Conclusion: based on the information provided, the investigation concluded that the most likely cause for the discrepant meca results were sensitivity/specificity differences between the filmarray bcid panel and the comparator methods with a polymicrobial specimen.Since the filmarray bcid panel detects organisms directly from blood culture bottles rather than isolated colonies, discrepancies with standard identification techniques can occur and are a part of normal system performance.The filmarray bcid panel run files suggest that this may have been a polymicrobial specimen, where the bottle could have been driven to positivity by a high titer s.Aureus which could have outgrown another staphylococcus species containing the (b)(6).It is also possible that there was a heterogeneous population of s.Aureus, where a small portion of the population carried the (b)(6).The staphylococcus species containing the (b)(6) may not have been at a titer that was detectable by culture.As stated in the filmarray bcid panel instruction booklet [rfit-prt-0369], the filmarray bcid panel may not distinguish mixed cultures when two or more species of the same genus or organism group are present in a specimen (e.G., s.Aureus and s.Epidermidis).According to table 26 of the filmarray bcid panel instruction booklet [rfit-prt-0369], the performance claim for the (b)(6) assay in staphylococcus, s.Aureus detected specimens compared to phenotypic ast methods showed a ppa of 97.7% (95% ci 93.4-99.5%) and an npa of 96.7% (95% ci 91.7-99.1%).Results from the filmarray bcid panel must be correlated with the clinical history, epidemiological data, and other data available to the clinician evaluating the patient.In mixed growth, the filmarray bcid panel does not specifically attribute meca-mediated methicillin resistance to either s.Aureus or other staphylococcus sp.
 
Event Description
Summary: (b)(6) hospital and medical center ((b)(6)) reported discrepant meca results on the filmarray® blood culture identification (bcid) panel after testing a sample from a (b)(6)-year-old, female patient.Due to the filmarray bcid panel results, the patient was treated with vancomycin for 6 weeks.Later in the course of therapy the patient experienced a permanent kidney function impairment that was caused in part by the treatment of vancomycin.Upon investigation, no malfunction occurred and the system was working within specification.The investigation concluded that the most likely cause for the meca discrepant result was sensitivity/specificity differences between the filmarray bcid panel and comparator methods with a polymicrobial specimen.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FILMARRAY® BLOOD CULTURE IDENTIFICATION PANEL
Type of Device
FILMARRAY® BLOOD CULTURE IDENTIFICATION PANEL
Manufacturer (Section D)
BIOFIRE DIAGNOSTICS, LLC
515 colorow drive
salt lake city, ut
Manufacturer (Section G)
BIOFIRE DIAGNOSTICS, LLC
515 colorow drive
salt lake city, ut
Manufacturer Contact
kristen kanack
515 colorow drive
salt lake city, ut 
7366354133
MDR Report Key10080024
MDR Text Key195002449
Report Number3002773840-2020-00002
Device Sequence Number1
Product Code PEN
UDI-Device Identifier00815381020086
UDI-Public00815381020086
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K181493
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator No Information
Device Expiration Date11/04/2020
Device Model NumberRFIT-ASY-0126
Device Catalogue NumberRFIT-ASY-0126
Device Lot Number1190119
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/23/2020
Initial Date FDA Received05/21/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/12/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age73 YR
-
-