Model Number 0570-0313 |
Device Problems
Poor Quality Image (1408); Material Deformation (2976)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 04/04/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Upon review of the device history for the serial number (b)(4) it was determined that the device was manufactured on 30 oct 2017 and is past the two (2) year expected product life as outlined in the glidescope operations and maintenance manual (omm).Due to the age of the device and the fact that there are no repairs available for the video baton the customer was advised to replace their glidescope avl video baton 3-4.The customer declined to have their video baton returned for evaluation and disposal.At this time, the cause of the reported issue could not be determined; however, it is likely that the age of the device, 2.6 years, caused or contributed to the event.Corrective action is not required at this time.Verathon will continue to monitor for trends.
|
|
Event Description
|
The customer reported that during a patient procedure, using a glidescope avl video baton 3-4, the image was intermittent during intubation.The biomed reported that the cable felt kinked at the base of the baton.No delay in the procedure, use of a backup device, or harm to the patient or user was reported.
|
|