Model Number 96102
Device Problem Loss of Power (1475)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/27/2020
Event Type  malfunction  
Manufacturer Narrative
Spacelabs technical support investigation found a third-party ups serving the telemetry xtr system had failed causing the reported problem.The reported problem was verified and customer moved everything over to a new ups, to resolve the issue.Normal device operation was verified and the device was restored to service.This investigation is considered complete and the issue closed.
 
Event Description
Customer reported that on (b)(6) 2020 that the hospitals lost all telemetry beds.They show as offline.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPACELABS XHIBIT CENTRAL STATION
Type of Device
XHIBIT CENTRAL MONITOR
Manufacturer (Section D)
SPACELABS HEALTHCARE INC.
35301 se center st.
snoqualmie WA 98065
Manufacturer Contact
marc rivas
35301 se center st.
snoqualmie, WA 98065
4253635568
MDR Report Key10097106
MDR Text Key204697556
Report Number3010157426-2020-00005
Device Sequence Number1
Product Code MHX
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K122146
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Remedial Action Repair
Type of Report Initial
Report Date 01/01/2005,05/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number96102
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Initial Date Manufacturer Received 04/27/2020
Initial Date FDA Received05/28/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-