The patient was undergoing a thrombectomy procedure in the m1 segment of the middle cerebral artery (mca) using a penumbra system jet 7 reperfusion catheter (jet7), guidewire, and non-penumbra microcatheter.It was reported that the patient¿s anatomy was tortuous.During the procedure, the physician placed the jet7, guidewire, and microcatheter in the target vessel and aspirated using the jet7.As the aspiration stopped, the physician removed the jet7.Upon removal, the jet7 was found ovalized four to five centimeters from the distal end; therefore, the jet7 was not used for the remainder of the procedure.It was also reported that without aspiration pressure, the jet7 reverted to its original form.The procedure was completed using another jet7 and the same guidewire and microcatheter.There was no report of an adverse effect to the patient.
|
Results: the jet7 was kinked approximately 29.5 cm from the hub.The jet7 was stretched approximately 127.0-132.5 cm from the hub.Stretch marks were present approximately 122.0-123.0 cm from the hub.Conclusions: evaluation of the returned jet7 revealed that the distal shaft was stretched.If the jet7 is forcefully retracted during use, damage such as stretching may occur.During functional testing, the jet7 was connected to demonstration pump and able to aspirate fluid without an issue.There was some deformation noted under aspiration due to stretch damage.No other devices associated with the complaint were returned for evaluation.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|