SMITH & NEPHEW, INC. LARGE SCREWDRIVER HANDLE; INSTRUMENT, MANUAL, SURGICAL, GENERAL USE Back to Search Results
Catalog Number 71173547
Device Problems Difficult to Remove (1528); Connection Problem (2900)
Patient Problem No Patient Involvement (2645)
Event Date 05/01/2020
Event Type  malfunction  
Event Description
It was reported that during inspection the driver was found to be stuck in handle.No case related.No injury, no backup needed.
 
Manufacturer Narrative
The devices, intended use in treatment, were returned for evaluation.A visual inspection confirmed the hexdriver is stuck in the screwdriver handle.The devices show significant signs of wear/usage.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A review of complaint history for the listed part revealed no prior complaints for the listed batch.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.This is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LARGE SCREWDRIVER HANDLE
Type of Device
INSTRUMENT, MANUAL, SURGICAL, GENERAL USE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10099822
MDR Text Key193392703
Report Number1020279-2020-01897
Device Sequence Number1
Product Code MDM
UDI-Device Identifier03596010519580
UDI-Public3596010519580
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 09/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number71173547
Device Lot Number18LTM1146
Initial Date Manufacturer Received 05/01/2020
Initial Date FDA Received05/29/2020
Supplement Dates Manufacturer Received09/07/2020
Supplement Dates FDA Received09/21/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-