SMITH & NEPHEW, INC. LARGE SCREWDRIVER HANDLE; INSTRUMENT, MANUAL, SURGICAL, GENERAL USE
|
Back to Search Results |
|
Catalog Number 71173547 |
Device Problems
Difficult to Remove (1528); Connection Problem (2900)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 05/01/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that during inspection the driver was found to be stuck in handle.No case related.No injury, no backup needed.
|
|
Manufacturer Narrative
|
The devices, intended use in treatment, were returned for evaluation.A visual inspection confirmed the hexdriver is stuck in the screwdriver handle.The devices show significant signs of wear/usage.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A review of complaint history for the listed part revealed no prior complaints for the listed batch.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.This is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.
|
|