Patient information was not provided.Cardiac-assist inc.Manufactures the protek duo veno-venous cannula.The incident occurred in (b)(6).A picture of the cannula was provided and analyzed.The cannula appeared to be completely fractured at the distal side of the transition region.The wire wrapping is exposed and the breach extends to the lumen.The cannula fracture is hypothesized to be due to placement of the transition region in the right ventricle.The configuration places the cannula in a "u" shape and therefore the damage at the transition region is likely due to cyclic loading from repeated right ventricle contraction during support.In this case, the cannula kink was not detected until the explant and did not impact the use of the device.Based on the current level of information, it cannot be excluded that the reported event may have contributed to the reported hemolysis.The instruction for use, 7000-0155 rev 10, contains a warning regarding kinking.Section 2, warnings and precautions item 13 states the following: "avoid kinking the protek-duo cannula or introducer during insertion and placement within the patient's vasculature.Note: the area of the cannula that tapers from large to small diameter (transition region) is more susceptible to kinking than other areas.Avoid positioning this region with a tight bend radius during support".
|