Model Number NK652D
Device Problem Fracture (1260)
Patient Problem Patient Problem/Medical Problem (2688)
Event Date 05/09/2020
Event Type  Injury  
Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Additional information / investigation results will be provided in a supplemental report.
 
Event Description
It was reported that there was an issue with product biolox delta prosth.Head 12/14 36 mm l.According to complaint description it was reported the following: the patient, during normal daily activities, suddenly had shooting pain, feeling of instability in the left leg and inability to walk.After 9 year, 9 month there was postoperative (po) fracture of ceramic ball head.A revision surgery was necessary.Additional information was not provided nor available.The adverse event is filed under aag reference (b)(6).Involved components: nk652d - biolox delta prosth.Head 12/14 36 mm l - batch 51672173.Nh109d - sc/msc biolox delta ins.36 mm 56/58 sym.- batch 51676875.Nc286t - metha cap 12/14 135°/0° size 6 - batch 51472758.Nh056t - plasmacup sc size 56 mm - batch 51659564.
 
Manufacturer Narrative
A - patient data updated.General information: according to the provided information, we will receive the components for investigation.They will be forwarded to the manufacturer of the ceramic components (ceramtec gmbh, plochingen).Afterwards this report will be updated after the investigation has been executed.Involved components: nk652d - biolox delta prosth.Head 12/14 36mm l - batch 51672173.Nh109d - sc/msc biolox delta ins.36mm 56/58 sym.- batch 51676875.Nc286t - metha cap 12/14 135°/0° size 6 - batch 51472758.Nh056t - plasmacup sc size 56mm - batch 51659564.Consequences for the patient: post-operative medical intervention was necessary - revision surgery.Investigation: investigation ongoing.Pictorial documentation: investigation ongoing.Batch history review: the device history records have been checked for the available lot numbers and found to be according to the specification, valid at the time of production.No further complaints registered against the same lot number (51672173).Conclusion and root cause: to date, we suspect that the failure is most probably usage or patient related.There is no indication for a material or manufacturing failure.However, the investigation is still ongoing.This report will be updated after the investigation has been executed.Rationale: refer to conclusion and root cause.
 
Manufacturer Narrative
The device quality and manufacturing history records (dhr) have been checked for all available lot numbers and the products found to be according to our specification valid at the time of production.Explanation and rationale.· the density of the ball head was analysed and found to be complying with the delivery specification for biolox® components.The microstructure as obtained from the quality documents fulfils the requirements as specified at the time of production.There are no indications of any pre-existing material defect.· secondary metal transfer patterns can be found on the fragments of the ball head as a result of contact with metal parts and/ or surgical instruments after the fracture event.· the expected primary metal transfer can be found on the cone of the ball head.Additionally, metal transfer can be found in region d/e.However, it cannot be ascertained whether this metal transfer occurred prior to or after the primary fracture event.Parts of the primary fracture surfaces and the region of the fracture origin can be identified.Due to secondary damages the fracture origin cannot be determined on the fragments of the ball head.· due to missing fragments and secondary damages, it cannot be drawn any conclusion regarding a possible cause for the fracture of the ball head.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOLOX DELTA PROSTH.HEAD 12/14 36MM L
Type of Device
HIP ENDOPROSTHESES - HEADS
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
MDR Report Key10101471
MDR Text Key193654450
Report Number9610612-2020-00179
Device Sequence Number1
Product Code LZO
Combination Product (y/n)N
PMA/PMN Number
K082991
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup,Followup
Report Date 10/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2020
Device Model NumberNK652D
Device Catalogue NumberNK652D
Device Lot Number51672173
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 05/14/2020
Initial Date FDA Received05/29/2020
Supplement Dates Manufacturer Received06/05/2020
09/07/2020
Supplement Dates FDA Received06/30/2020
10/05/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
NC286T - BATCH 51472758.; NH056T - BATCH 51659564.; NH109D - BATCH 51676875.; NK652D - BATCH 51672173
Patient Outcome(s) Required Intervention;
Patient Age38 YR
-
-