At this time, the suspect device has not been returned for evaluation.However, vyaire technical support suggested customer to calibrate r17 which calibrates the panel meter.The customer reported that the panel meter is back to normal by adjusting r17.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
Device evaluation: d10, g4, g7, h2, h3, h6 and h10.Result of investigation: the vyaire failure analysis laboratory received the unit.A visual inspection of driver assembly was performed and no visual sign of physical damage was found.The printed circuit board assembly was installed into a known good top level unit and measured test point tp8 and powered on.The frequency panel meter was showing 15hz and the oscilloscope is reading 14.7hz (reading good).Performed the 6k preventive maintenance and the unit passed.The unit was then left to cycle for 2 hours with no fluctuations in frequency readings.
|