• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE MEDICAL 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY Back to Search Results
Model Number 3100 A
Device Problem Mechanical Problem (1384)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/05/2020
Event Type  Injury  
Manufacturer Narrative
At this time, the suspect device has not been returned for evaluation.However, vyaire technical support suggested customer to calibrate r17 which calibrates the panel meter.The customer reported that the panel meter is back to normal by adjusting r17.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event Description
The customer reported that the frequency of 3100a was drifting from 15 to 10.The issue occurred during patient-use and the device was replaced with another ventilator.The customer confirmed that there was no patient harm associated with the reported event.
 
Manufacturer Narrative
Device evaluation: d10, g4, g7, h2, h3, h6 and h10.Result of investigation: the vyaire failure analysis laboratory received the unit.A visual inspection of driver assembly was performed and no visual sign of physical damage was found.The printed circuit board assembly was installed into a known good top level unit and measured test point tp8 and powered on.The frequency panel meter was showing 15hz and the oscilloscope is reading 14.7hz (reading good).Performed the 6k preventive maintenance and the unit passed.The unit was then left to cycle for 2 hours with no fluctuations in frequency readings.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV)
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
VYAIRE MEDICAL
26125 n. riverwoods blvd.
mettawa, il
MDR Report Key10107158
MDR Text Key193282912
Report Number2021710-2020-12047
Device Sequence Number1
Product Code LSZ
UDI-Device Identifier10846446003079
UDI-Public(01)10846446003079(11)20111201
Combination Product (y/n)N
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 05/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100 A
Device Catalogue Number768901
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/14/2020
Initial Date Manufacturer Received 05/05/2020
Initial Date FDA Received06/01/2020
Supplement Dates Manufacturer Received08/14/2020
Supplement Dates FDA Received09/22/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-