SMITH & NEPHEW, INC. RDCE FRCPS W/ PTS BRD; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
|
Back to Search Results |
|
Catalog Number 71173377 |
Device Problem
No Apparent Adverse Event (3189)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 05/05/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that when the evos small instrument tray got opened for a case at parmc it was noticed that the clamps were missing.The case had not been started yet so no patient or doctor was involved.The tips are just bent.
|
|
Manufacturer Narrative
|
The devices, used in treatment, were returned for evaluation.A visual inspection confirmed the rdce frcps w/ pts brd are bent.The devices show significant signs of wear/usage.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.This is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.
|
|