• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. RDCE FRCPS W/ PTS BRD; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. RDCE FRCPS W/ PTS BRD; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Catalog Number 71173377
Device Problem Deformation Due to Compressive Stress (2889)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/07/2020
Event Type  malfunction  
Event Description
It was reported that before procedure the rdce frcps w/ pts brd were found bent.
 
Manufacturer Narrative
The device, intended for use in treatment, was returned for evaluation.The visual inspection of the returned forcep confirms the tip is bent.This device was manufactured in 2020.This device shows signs of significant wear and usage.A review of complaint history on the listed part revealed no prior complaints for the listed batch with the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident this device is a reusable instruments that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RDCE FRCPS W/ PTS BRD
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10113785
MDR Text Key193591100
Report Number1020279-2020-01962
Device Sequence Number1
Product Code JDI
UDI-Device Identifier03596010519535
UDI-Public3596010519535
Combination Product (y/n)N
PMA/PMN Number
K123598
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 12/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71173377
Device Lot Number20CTCD034
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/09/2020
Initial Date Manufacturer Received 05/07/2020
Initial Date FDA Received06/03/2020
Supplement Dates Manufacturer Received11/25/2020
Supplement Dates FDA Received12/05/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-