Model Number DBL OFFSET BROACH HDLE RT
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/04/2020
Event Type  malfunction  
Event Description
It was reported that during and efip inspection the item did into work correctly.The dbl offset broach hdle right did not lock correctly.
 
Manufacturer Narrative
The device, intended for use in treatment, was returned for evaluation.A visual inspection of the returned double offset broach handle (right) confirms the handles exhibit significant signs of use and wear.There are nicks, scratches and gouges in this device.This device was manufactured in 2014.A functional evaluation was conducted and confirms the handle will not lock in place.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DBL OFFSET BROACH HDLE RT
Type of Device
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10117117
MDR Text Key193921191
Report Number1020279-2020-01973
Device Sequence Number1
Product Code JDG
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 09/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberDBL OFFSET BROACH HDLE RT
Device Catalogue Number71360090
Device Lot Number14FM05775
Was Device Available for Evaluation? Yes
Distributor Facility Aware Date05/04/2020
Initial Date Manufacturer Received 05/04/2020
Initial Date FDA Received06/03/2020
Supplement Dates Manufacturer Received09/06/2020
Supplement Dates FDA Received09/11/2020
Patient Sequence Number1
-
-