• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PINN ESC LINER EXTRACTOR TIP; HIP INSTRUMENTS : EXTRACTION INSTRUMENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PINN ESC LINER EXTRACTOR TIP; HIP INSTRUMENTS : EXTRACTION INSTRUMENTS Back to Search Results
Model Number 2218-00-010
Device Problems Material Twisted/Bent (2981); Scratched Material (3020)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/27/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the instrument is dented and deformed.No surgical delay.
 
Manufacturer Narrative
Product complaint # : (b)(4).Investigation summary: examination of the returned instrument confirmed the complaint.Depuy-synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.
 
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN ESC LINER EXTRACTOR TIP
Type of Device
HIP INSTRUMENTS : EXTRACTION INSTRUMENTS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10118721
MDR Text Key193982552
Report Number1818910-2020-13187
Device Sequence Number1
Product Code HWB
UDI-Device Identifier10603295099505
UDI-Public10603295099505
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 05/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2218-00-010
Device Catalogue Number221800010
Device Lot NumberNW239213
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/04/2020
Initial Date Manufacturer Received 05/27/2020
Initial Date FDA Received06/04/2020
Supplement Dates Manufacturer Received06/08/2020
06/24/2020
Supplement Dates FDA Received06/12/2020
06/25/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-