MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC AIR PEN DRIVE; MOTOR, DRILL, PNEUMATIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC AIR PEN DRIVE; MOTOR, DRILL, PNEUMATIC Back to Search Results
Catalog Number 05.001.080
Device Problems Output below Specifications (3004); Intermittent Loss of Power (4016)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/01/2020
Event Type  malfunction  
Manufacturer Narrative
Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.The complete address of the manufacturing location is currently not available.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
 
Event Description
It was reported from (b)(6) that the air pen drive device would not hold the power and was inconsistent.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure or if a spare device was available.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2020.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Device evaluation: the actual device was returned for evaluation.A visual and functional assessment was performed which determined that the motor had too little power, the device leaked, and the turning ring ran hard.During repair, an evaluation was performed, and it was determined that the device had a worn motor, an air leak, a jammed/seized knob, and a component damage.It was further observed that the device had an insufficient/low power and a cracked/damaged housing.It was further determined that the device failed pretest for check the adjustment sleeve, check power with power test bench, check speed with tachometer, check internal leak tightness and check external leak tightness.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to component failure from normal wear.If additional information should become available, a supplemental medwatch report will be submitted accordingly.Correction: d10: the date device returned to manufacturer was documented as june 9, 2020 in the initial report.This has been corrected to june 2, 2020.
 
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.D10, h3, h6: the actual device has been returned and is currently pending evaluation.Once the investigation has been completed, a supplemental medwatch report will be submitted accordingly.Correction: g1-2: the manufacturer location was incorrectly documented as depuy synthes supplier and the complete address of the manufacturing location was unknown in the initial report.The location has been updated to oberdorf.The contact office name/address has been updated accordingly to reflect the corrected manufacturing facility.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIR PEN DRIVE
Type of Device
MOTOR, DRILL, PNEUMATIC
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key10118724
MDR Text Key194020950
Report Number8030965-2020-03871
Device Sequence Number1
Product Code HBB
UDI-Device Identifier07611819830899
UDI-Public07611819830899
Combination Product (y/n)N
PMA/PMN Number
K093361
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 05/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05.001.080
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/02/2020
Initial Date Manufacturer Received 05/25/2020
Initial Date FDA Received06/04/2020
Supplement Dates Manufacturer Received06/08/2020
06/22/2020
Supplement Dates FDA Received06/16/2020
06/29/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-