It was reported to philips the heartstart mrx defibrillator was unable to defibrillate while attempting to resuscitate a patient in cardiac arrest.The monitor alerted ¿unable to deliver shock, check pad placement¿.The paramedic crew performed trouble shooting procedures and attempted again.The ¿unable to deliver shock, check pad placement¿ message appeared again.The patient was placed on another unit¿s monitor and transported to the nearest facility.While enroute, another attempt was made with the same monitor to cardiovert the patient and the ¿unable to deliver shock, check pad placement¿ was noted again.The device was reported to be in use on a patient, causing a delay in therapy/treatment and will be considered a serious injury.However, no direct adverse event to the patient or user was reported.A philips clinician reviewed the electronic event file from this event.The device was powered on into manual mode and pads were placed.During this event, the users attempted to deliver shocks 3 times at 200j and once at 100j in the sync mode.All of the shocks were aborted due to a patient impedance outside the range for the delivery of therapy.Of note, there were multiple pads off and on messages consistent with an insufficient pads to patient connection.The customer reported receiving messages directing them to check the pads.The heartstart mrx instructions for use troubleshooting section (publication number 453564307761, page 316) describes two messages that may appear related to pads and impedance: "no shock delivered, replace pads now" and "no shock delivered, press pads firmly".The user is directed to make sure that the pads are applied properly and to replace the pads if needed.A philips field service engineer evaluated the device and was unable to duplicate the issue.No parts were replaced.The device passed all performance assurance tests and was placed back into use with the customer.Based on the device testing and information in the event file, philips determined there was no malfunction of the device.
|