• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE MEDICAL 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY Back to Search Results
Model Number 3100 A
Device Problem Pressure Problem (3012)
Patient Problem No Patient Involvement (2645)
Event Date 05/11/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Vyaire field service representative (fsr) went onsite and evaluated the ventilator.Fsr was able to duplicate the issue of not pressurizing.Reseated the 3 cap/diaphragm and then the ventilator would pressurize.Fsr attempted to recreate the original issue, but could not.In case the issue is highly intermittent, all of the electrical connections under the top cover were reseated.Many more attempts were made to try to recreate the reported issue, but it could not be duplicated.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event Description
The customer reported device would not pressurize and start ventilation on the 3100 a ventilator.The customer confirmed that there was no patient involvement that was associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV)
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
VYAIRE MEDICAL
26125 n. riverwoods blvd.
mettawa, il
Manufacturer (Section G)
VYAIRE MEDICAL INC.
1100 bird center drive
palm springs, ca
Manufacturer Contact
mindy faber
26125 n. riverwoods blvd.
mettawa, il 
MDR Report Key10121634
MDR Text Key194119015
Report Number2021710-2020-12021
Device Sequence Number1
Product Code LSZ
UDI-Device Identifier10846446003079
UDI-Public(01)10846446003079(11)20051201
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100 A
Device Catalogue Number768901
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/11/2020
Initial Date FDA Received06/04/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-