Catalog Number 71440190
Device Problem Flaked (1246)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/08/2020
Event Type  malfunction  
Event Description
It was reported that, before surgery, the surgeon stated that the black plastic was shredding off.No delay to procedure, no injury to patient.A s&n backup was available.
 
Manufacturer Narrative
The device, used in treatment, was returned for evaluation.A visual inspection of the returned device confirms the black coating from the bumper is peeling off.The device was manufactured in 2019.The device shows signs of normal wear and use.A review of complaint history on the listed part revealed no prior complaint for the listed batch with the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII FEMORAL IMPACTOR
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10122539
MDR Text Key194115003
Report Number1020279-2020-01995
Device Sequence Number1
Product Code JDI
Combination Product (y/n)N
PMA/PMN Number
K123598
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 07/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71440190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/09/2020
Initial Date Manufacturer Received 05/08/2020
Initial Date FDA Received06/05/2020
Supplement Dates Manufacturer Received05/08/2020
07/06/2020
Supplement Dates FDA Received06/05/2020
07/15/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-