MAUDE Adverse Event Report: TORAX MEDICAL, INC. LINX 1.5 16 BEAD US; ANTI-REFLUX IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORAX MEDICAL, INC. LINX 1.5 16 BEAD US; ANTI-REFLUX IMPLANT Back to Search Results
Model Number LXMC16
Device Problem Patient Device Interaction Problem (4001)
Patient Problems Dysphagia/ Odynophagia (1815); No Code Available (3191)
Event Date 05/13/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4).Date sent: 6/5/2020.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The lot was not provided; therefore, the manufacturing record evaluation could not be performed.The following information was requested, but unavailable: prior to linx placement, did the patient have an egd, ph, and manometry studies done? if yes, could you please share the results? on what date did the implant take place? what is the lot number of the linx device? when using the linx sizing device what technique was used to determine the size? did the patient have an autoimmune disease? is the patient currently taking steroids / immunization drugs? did the patient have any pre-existing dysphagia or other conditions (other than gerd)? how severe was the dysphagia/odynophagia before intervention? were there any intra-operative complications during implant? was there any hiatal or crural repair done at the same time as the implant? were there any other contributing factors that led to the removal of the device other than the reported dysphagia? was the device found in the correct position/geometry at the time of removal?.
 
Event Description
It was reported that post implant of a linx device an explant of lxmc16 due to dysphagia was performed.There were no patient consequences.
 
Manufacturer Narrative
(b)(4).Date sent: 07/16/2020.Device analysis: overall review of the device function and dimensions show no anomalies from a device that has been reasonably changed as part of the explant procedure.Visual analysis was consistent with an explanted device, and link length and tensile force were found to meet the applicable specifications.Overall, no analysis conclusions relevant to the patient experience were found.The lot/batch was not provided; therefore, the manufacturing records could not be reviewed.
 
Manufacturer Narrative
(b)(4).Date sent: 07/10/2020.Additional information was requested, and the following was received: no further information on this compliant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINX 1.5 16 BEAD US
Type of Device
ANTI-REFLUX IMPLANT
Manufacturer (Section D)
TORAX MEDICAL, INC.
4188 lexington avenue north
shoreview MN
MDR Report Key10122861
MDR Text Key194779527
Report Number3008766073-2020-00073
Device Sequence Number1
Product Code LEI
UDI-Device Identifier00855106005363
UDI-Public00855106005363
Combination Product (y/n)N
PMA/PMN Number
P100049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup,Followup
Report Date 05/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberLXMC16
Device Catalogue NumberLXMC16
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer06/18/2020
Initial Date Manufacturer Received 05/14/2020
Initial Date FDA Received06/05/2020
Supplement Dates Manufacturer Received06/15/2020
07/14/2020
Supplement Dates FDA Received07/10/2020
07/16/2020
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-