• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO BCT TRIMA ACCEL; TRIMA ACCEL AUTOMATED BLOOD COLLECTION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO BCT TRIMA ACCEL; TRIMA ACCEL AUTOMATED BLOOD COLLECTION SYSTEM Back to Search Results
Model Number 81000
Device Problems Mechanical Problem (1384); Device Slipped (1584)
Patient Problem No Patient Involvement (2645)
Event Date 05/19/2020
Event Type  malfunction  
Manufacturer Narrative
Investigation: a terumo bct service technician checked out the machine at the customer site and was able to confirm the reported condition.The screw from the iv pole release button was making too much contact with the releases latch.The screw was adjusted.The service representative verified the iv pole was functioning as expected without sliding down.The device serial number history report indicates no further related issues have been reported for this device.One year of service history was reviewed for this device with no issues related to the reported condition identified.Corrective action: an internal capa has been initiated to evaluate reports of the iv pole dropping down suddenly.Root cause: the root cause of this failure was the iv pole adjuster button screw was making too much contact with iv pole latch.
 
Event Description
The customer reported that the iv pole slowly works its way down during the procedure, the iv pole has not fallen and no injuries have occurred as a result of the iv pole problem.The iv pole slide guard is in place on this machine.No injury was reported for this incident and no patient was connected at the time the iv pole was sliding down, therefore no patient information is reasonably known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIMA ACCEL
Type of Device
TRIMA ACCEL AUTOMATED BLOOD COLLECTION SYSTEM
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
Manufacturer Contact
scot hilden
10810 w collins ave
lakewood, CO 80215
3032314970
MDR Report Key10125404
MDR Text Key200866068
Report Number1722028-2020-00280
Device Sequence Number1
Product Code GKT
UDI-Device Identifier05020583810006
UDI-Public05020583810006
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
BK180231
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number81000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 05/19/2020
Initial Date FDA Received06/05/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/14/2018
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-