Catalog Number 103540000
Device Problems Fracture (1260); Off-Label Use (1494); Naturally Worn (2988)
Patient Problem No Code Available (3191)
Event Date 05/26/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that a wright medical head was used with our bipolar head, causing suspicion poly wear and instability.Doi: unknown; dor: (b)(6) 2020; affected side: right hip.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : examination of the returned device finds no obvious poly material wear or unusual wear pattern.It is noted that the head is competitor manufactured.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELF CENT HIP 40X28 RUST
Type of Device
HEMI HIP IMPLANT : HIP ACETABULAR POLY/METAL HEMI
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10130855
MDR Text Key194457060
Report Number1818910-2020-13406
Device Sequence Number1
Product Code KWY
Combination Product (y/n)N
PMA/PMN Number
K812672
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 05/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number103540000
Device Lot Number578696
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/04/2020
Initial Date Manufacturer Received 05/26/2020
Initial Date FDA Received06/08/2020
Supplement Dates Manufacturer Received06/04/2020
06/25/2020
Supplement Dates FDA Received06/11/2020
06/28/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
HIP FEMORAL HEAD (COMPETITOR); UNKNOWN HIP ACETABULAR POLY/METAL HEMI
Patient Outcome(s) Required Intervention;
-
-