VYAIRE MEDICAL 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY Back to Search Results
Model Number 3100 A
Device Problem Mechanical Problem (1384)
Patient Problem No Patient Involvement (2645)
Event Date 05/12/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Any additional information provided by the customer will be included in a follow up report.Vyaire field service representative (fsr) went onsite and evaluated the ventilator.It was confirmed that the bias flow would not go above 15.Fsr bypassed blender and issue persisted.Opened cover and found clear tubing for bias flow crimped.Removed crimp and bias flow returned to its normal state.Calibrated bias flow solenoid to max at 40.Ran performance test.Unit passed all test and runs according to manufacturer's specifications.
 
Event Description
The customer reported that the 3100a's bias flow meter will not rise above 15lpm.The customer confirmed that there was no patient involvement associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV)
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
VYAIRE MEDICAL
26125 n. riverwoods blvd.
mettawa, il
Manufacturer (Section G)
VYAIRE MEDICAL INC.
1100 bird center drive
palm springs, ca
Manufacturer Contact
mindy faber
26125 n. riverwoods blvd.
mettawa, il 
MDR Report Key10131368
MDR Text Key194532987
Report Number2021710-2020-12046
Device Sequence Number1
Product Code LSZ
UDI-Device Identifier10846446003079
UDI-Public(01)10846446003079(11)20071116
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100 A
Device Catalogue Number768901
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/12/2020
Initial Date FDA Received06/08/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/16/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-