• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC STEALTHSTATION S8 EM ENT SYSTEM; INSTRUMENT, STEREOTAXIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC STEALTHSTATION S8 EM ENT SYSTEM; INSTRUMENT, STEREOTAXIC Back to Search Results
Model Number 9735669
Device Problem Output Problem (3005)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/03/2020
Event Type  malfunction  
Manufacturer Narrative
Other relevant device(s) are: product id: 9735856, serial/lot: unknown.Manufacturer representative resolved the issue after the procedure.No system checkout necessary.Patient information requested.Surgeon information requested.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a navigation system being used during a procedure.It was reported that the monitor had a slight flicker.The surgeon opted to continue to use the system for the remainder of the procedure as tracking was functional and the flicker was minimal.After the case, the manufacturer representative reseated the monitor cabling and the issue was resolved.No patient harm reported.
 
Manufacturer Narrative
H2) additional information: see b5 and e1.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received additional information that the type of procedure was functional endoscopic sinus surgery (fess).There was no reported delay or impact to patient outcome due this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S8 EM ENT SYSTEM
Type of Device
INSTRUMENT, STEREOTAXIC
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
MDR Report Key10132779
MDR Text Key194493382
Report Number1723170-2020-01614
Device Sequence Number1
Product Code PGW
UDI-Device Identifier00643169838918
UDI-Public00643169838918
Combination Product (y/n)N
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 07/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9735669
Device Catalogue Number9735669
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 06/03/2020
Initial Date FDA Received06/09/2020
Supplement Dates Manufacturer Received07/09/2020
Supplement Dates FDA Received07/28/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-