ZIMMER BIOMET, INC. TPRLC XR T1 PPS 14X148MM; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Device Contaminated During Manufacture or Shipping (2969)
Patient Problem No Patient Involvement (2645)
Event Date 05/28/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Foreign report source: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034 - 2020 - 02281, 0001825034 - 2020 - 02282, 0001825034 - 2020 - 02283, 0001825034 - 2020 - 02284, 0001825034 - 2020 - 02287.
 
Event Description
It was reported that during an investigation of circulated items, several devices were identified as having debris in their sterile package.No patients were involved.No additional information is available.
 
Event Description
Upon receipt of additional information, it has been determined that the product is conforming to specifications.Event is no longer considered reportable, and the initial report was forwarded in error and should be voided.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.Updated: b4, b5, d4, d10, g4, h2, h3, h4, h6, h10.Complaint sample was evaluated and the reported event was confirmed.Debris inside the sterile packaging is consistent with the appearance of the porous coating and foam debris from the foam packaging inside the sterile barrier.Dhr was reviewed and no discrepancies were found.The likely condition of the product when it left zimmer biomet was conforming to specification.The root cause of the reported event is likely to be due to transit damage causing the foam packaging to become abraded and shed.D4: udi:(b)(4).Upon investigation, it has been determined that the debris in the sterile packaging meets the acceptable criteria and product is conforming to specifications.Event is no longer considered reportable, and initial report should be voided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC XR T1 PPS 14X148MM
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key10134423
MDR Text Key194673296
Report Number0001825034-2020-02285
Device Sequence Number1
Product Code LPH
Combination Product (y/n)N
PMA/PMN Number
K120030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number51-105140
Device Lot Number3786675
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/07/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/28/2020
Initial Date FDA Received06/09/2020
Supplement Dates Manufacturer Received08/31/2020
Supplement Dates FDA Received09/04/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-