• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL PUERTO RICO, INC. SJM REGENT HEART VALVE W/FLEX CUFF; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL PUERTO RICO, INC. SJM REGENT HEART VALVE W/FLEX CUFF; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 21AGFN-756
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/18/2020
Event Type  malfunction  
Manufacturer Narrative
Further information regarding this event has been requested.If returned, investigation results will be provided in a subsequent submission.
 
Event Description
On (b)(6) 2020, a sjm regent heart valve w/flex cuff was chosen for procedure.During the procedure, as the user pushed the aortic valve down into the annulus, one of the leaflets broke off the valve.
 
Manufacturer Narrative
The reported event of one of the leaflets breaking off the valve was confirmed.One leaflet was dislodged from the orifice and returned with the valve.A small chip was noticed in on recessed pivot area.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the leaflet dislodgment could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM REGENT HEART VALVE W/FLEX CUFF
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ST. JUDE MEDICAL PUERTO RICO, INC.
p.o. box 998
lot 20 b st.
caguas, puerto rico 00725
MDR Report Key10134461
MDR Text Key194520830
Report Number2648612-2020-00060
Device Sequence Number1
Product Code LWQ
UDI-Device Identifier05414734005975
UDI-Public05414734005975
Combination Product (y/n)N
PMA/PMN Number
P810002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Remedial Action Replace
Type of Report Initial,Followup
Report Date 07/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number21AGFN-756
Device Catalogue Number21AGFN-756
Device Lot Number7262621
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/18/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/20/2020
Initial Date FDA Received06/09/2020
Supplement Dates Manufacturer Received07/03/2020
Supplement Dates FDA Received07/20/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-