Investigation: the device history record was reviewed for this lot.There were no issues noted that would have contributed to this incident.The customer returned the used astotube blood warmer tubing for investigation.The customer did not return the spectra optia set with the tubing.Visual inspection confirmed the presence of sections of air, measuring between approximately 3-40 mm in length interspersed periodically throughout the blood in the tubing.No leaks, kinks/occlusions or manufacturing defects were identified.Investigation is in process.A follow up report will be provided.
|
This report is being filed to provide additional information in a.1 and h10.Investigation: the run data file (rdf) was analyzed for this event.Per review of the rdf, the rbcx procedure was completed successfully with no return-line air detector (rlad) alarms indicating that air in the blood warmer tubing was generated downstream of the rlad, which is consistent with outgassing of blood.Root cause: based on the investigation, the root cause of the air bubbles noted in blood warmer tubing may be related to outgassing.During exchange procedures on spectra optia, the replacement fluids may be cold.If they are not allowed to warm to room temperature, and if the return blood is warmed, air bubbles may form.The reason for this "outgassing" is that gasses are more soluble in liquids at low temperatures than at higher temperatures.If at a low storage temperature air is available to dissolve in a fluid and approaches its equilibrium solubility at that temperature, when the fluid is warmed the air will come out of solution, because its solubility is exceeded at the higher temperature.It is typically described as chains of very small bubbles or foam, which tend to rise toward the top of the tubing.The small bubbles may coalesce to form larger bubbles.It is also possible that the bubbles are the result of a small, unidentified luer leak at the blood warmer via a loose luer connection that was hung 50 cm or higher above the return access.
|