MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA STEM: QUADRA-H CEMENTLESS, HA COATED STD STEM SIZE 5

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA STEM: QUADRA-H CEMENTLESS, HA COATED STD STEM SIZE 5 Back to Search Results
Model Number 01.12.025
Device Problem Device Slipped (1584)
Patient Problem Hip Fracture (2349)
Event Date 05/13/2020
Event Type  Injury  
Manufacturer Narrative
Batch review performed by medacta regulatory affairs department on 19 may 2020: lot 1900008: (b)(4) items manufactured and released on 29-apr-2019.Expiration date: 14.04.2024.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.
 
Event Description
The patient dislocated in the hospital and after further investigation with x-ray it was discovered subsidence of the femoral component, the surgeon suspects that during the primary surgery 8 days before occurred a bone fracture, but it is not confirmed if the fracture occurred during of after the primary surgery.The surgeon revised the stem and head and cabled the fracture.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEM: QUADRA-H CEMENTLESS, HA COATED STD STEM SIZE 5
Type of Device
CEMENTLESS, HA COATED STD STEM
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
CH  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ   6874
Manufacturer Contact
stefano baj
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key10147233
MDR Text Key195584310
Report Number3005180920-2020-00326
Device Sequence Number1
Product Code JDI
UDI-Device Identifier07630030802065
UDI-Public07630030802065
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K082792
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 06/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date04/14/2024
Device Model Number01.12.025
Device Catalogue Number01.12.025
Device Lot Number1900008
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 05/13/2020
Initial Date FDA Received06/12/2020
Was Device Evaluated by Manufacturer? No
Date Device Manufactured04/29/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age71 YR
-
-