One swan ganz catheter with attached monoject 1.5 cc limited volume syringe was returned for evaluation.Pa distal lumen hub was completely detached from the extension tube.Indication of bonding was observed from cross-surface at hub.All through lumens were patent without any leakage or occlusion.The balloon inflated clear and concentric and remained inflated for 5 minutes without leakage.No visible damage or abnormality was observed from balloon or returned syringe.Balloon inflation test was performed using returned syringe with 1.5 cc air by holding the balloon under water for 5 minutes.Visual examinations were performed under microscope at 20x magnification.Customer report of ¿pa distal lumen hub was broken¿ was confirmed.An investigation has been initiated to consider any potential manufacturing factors that may have contributed to this complaint and implement any necessary corrective actions.Invasive procedures inherently involve some patient risks.Although serious complications associated with pulmonary artery catheters are relatively uncommon, the physician is advised before deciding to use the catheter to consider and weigh the potential benefits and risks associated with the use of the catheter against alternative procedures.The general risks and complications associated with indwelling catheters are well documented in the literature.It is standard clinical practice to inspect these devices prior to use on a patient.If a lumen disconnects from the hub during use, there would be risk of blood loss.The swan ganz catheter can be exchanged with a minimal delay in patient care and monitoring.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
|