This supplemental report is being submitted to provide the subject device evaluation result.The subject device was returned to olympus medical systems corp.(omsc) for evaluation.In the evaluation, it was found that the diaphragm pump was operating normally, but detergent residue was on the propeller in the sensor and was stuck.Omsc reviewed the manufacturing history record (dhr) of the subject device and confirmed no irregularity.The sensor had been sticking with the residue of the detergent, therefore the sensor did not function.The residue of the detergent was removed, then the sensor functioned properly.It was considered that this phenomenon was attributed to the sticking of the residue of the detergent due to long time disuse.Olympus stated the appropriate maintenance of oer-4 and the counter measures against abnormalities in the instruction manual of oer-4.
|