Catalog Number 72200568SR
Device Problem Defective Component (2292)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(6).
 
Event Description
It was reported that xenon light source's fan was not working.No case reported; therefore, there was no patient involvement.Results of investigation revealed an e12 temperature error which makes it a reportable event.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H11: corrected information on b5 (describe event or problem).H10: the reported device, intended for use in treatment, was received for evaluation.A visual inspection was performed on the exterior of product and no physical damage was observed.There was no relationship found between the returned device and the reported incident.The complaint of damaged/broken component could not be confirmed.Light source did fail for e-12 errors caused by an electronic component failure.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review found related failures; this failure mode will be trended to assess for any necessary corrective actions.No containment or corrective actions are recommended at this time.
 
Event Description
It was reported that xenon light source's fan was not working.No case reported; therefore, there was no patient involvement.Results of investigation revealed an e12 temperature error which makes it a reportable event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SVCE REPL {} LIGHT SCE {} 500XL {} XENON
Type of Device
IMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
MDR Report Key10153576
MDR Text Key195527883
Report Number1643264-2020-00384
Device Sequence Number1
Product Code FFS
Combination Product (y/n)N
PMA/PMN Number
K994084
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 07/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72200568SR
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/20/2020
Initial Date Manufacturer Received 04/06/2020
Initial Date FDA Received06/15/2020
Supplement Dates Manufacturer Received07/06/2020
Supplement Dates FDA Received07/07/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-