SMITH & NEPHEW, INC. SVCE REPL {} LIGHT SCE {} 500XL {} XENON; IMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE
|
Back to Search Results |
|
Catalog Number 72200568SR |
Device Problem
Defective Component (2292)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).
|
|
Event Description
|
It was reported that xenon light source's fan was not working.No case reported; therefore, there was no patient involvement.Results of investigation revealed an e12 temperature error which makes it a reportable event.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
Manufacturer Narrative
|
H11: corrected information on b5 (describe event or problem).H10: the reported device, intended for use in treatment, was received for evaluation.A visual inspection was performed on the exterior of product and no physical damage was observed.There was no relationship found between the returned device and the reported incident.The complaint of damaged/broken component could not be confirmed.Light source did fail for e-12 errors caused by an electronic component failure.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review found related failures; this failure mode will be trended to assess for any necessary corrective actions.No containment or corrective actions are recommended at this time.
|
|
Event Description
|
It was reported that xenon light source's fan was not working.No case reported; therefore, there was no patient involvement.Results of investigation revealed an e12 temperature error which makes it a reportable event.
|
|
Search Alerts/Recalls
|
|
|