Catalog Number 03666794001 |
Device Problem
Image Display Error/Artifact (1304)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 05/21/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The meter was requested for investigation.
|
|
Event Description
|
The initial reporter complained of a display issue with coaguchek xs meter serial number (b)(4).After replacing the batteries, the customer noticed segments missing in the results field of the device.Upon performing a display check, all results segments were visible.The customer stated there was a crack on the upper casing of the meter due to dropping it.The crack was not on the display screen.
|
|
Manufacturer Narrative
|
The customer's meter was received for investigation.The display showed no issues during the investigation.The meter¿s circuit board was tested for damage or contamination and it was found the printed circuit board (pcb) was contaminated.The observed contamination can temporarily lead to the reported display issue.The root cause is determined to be contamination of the contacts due to improper handling or maintenance by the customer.Medwatch fields d10 and h3 were updated.
|
|