• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COAGUCHEK XS SYSTEM; PROTHROMBIN TIME MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COAGUCHEK XS SYSTEM; PROTHROMBIN TIME MONITOR Back to Search Results
Catalog Number 03666794001
Device Problem Image Display Error/Artifact (1304)
Patient Problem No Patient Involvement (2645)
Event Date 05/21/2020
Event Type  malfunction  
Manufacturer Narrative
The meter was requested for investigation.
 
Event Description
The initial reporter complained of a display issue with coaguchek xs meter serial number (b)(4).After replacing the batteries, the customer noticed segments missing in the results field of the device.Upon performing a display check, all results segments were visible.The customer stated there was a crack on the upper casing of the meter due to dropping it.The crack was not on the display screen.
 
Manufacturer Narrative
The customer's meter was received for investigation.The display showed no issues during the investigation.The meter¿s circuit board was tested for damage or contamination and it was found the printed circuit board (pcb) was contaminated.The observed contamination can temporarily lead to the reported display issue.The root cause is determined to be contamination of the contacts due to improper handling or maintenance by the customer.Medwatch fields d10 and h3 were updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK XS SYSTEM
Type of Device
PROTHROMBIN TIME MONITOR
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
MDR Report Key10153594
MDR Text Key195119605
Report Number1823260-2020-01445
Device Sequence Number1
Product Code GJS
Combination Product (y/n)N
PMA/PMN Number
K062925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 07/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number03666794001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/23/2020
Initial Date Manufacturer Received 05/22/2020
Initial Date FDA Received06/15/2020
Supplement Dates Manufacturer Received05/22/2020
Supplement Dates FDA Received07/09/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-