Model Number 71420566
Device Problems Use of Device Problem (1670); Defective Device (2588)
Patient Problems Injury (2348); Joint Dislocation (2374)
Event Date 05/20/2020
Event Type  Injury  
Event Description
It was reported that after primary knee surgery, the patient experienced pain and dislocation.Revision surgery was performed.The surgeon indicated that the devices were not defective but there was a problem when the bones were shaved.It was too shallow to implant it.A backup device was used as replacement.Patient outcome is unknown.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
It was reported that after primary knee surgery, the patient experienced pain and dislocation.Revision surgery was performed.The affected genesis ii biconvex patellar component, used in treatment, was not returned for evaluation.Therefore a product analysis could not be performed.Our investigation including a review of manufacturing records did not reveal any deviation from the standard manufacturing processes.A review of the complaint history for the listed part revealed no prior complaints for the listed failure mode with the same batch number.There is no information that would suggest the device failed to meet specifications.A relationship, if any, between the device and the reported incident could not be corroborated.Based on this investigation, the need for corrective action is not indicated.The surgeon indicated that the devices were not defective but there was a problem when the bone was shaved.No medical documents were received for investigation.Thus no thorough investigation could not be performed.Some potential causes could include but are not limited to limited range of motion, surgical technique or patient anatomy.Without the return of the actual product involved and no patient medical records available, our investigation of this report is inconclusive.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.Should the device or additional information be received, the complaint will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GNS II BICONVEX PAT 23MM
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10155637
MDR Text Key195223615
Report Number1020279-2020-02406
Device Sequence Number1
Product Code HRY
UDI-Device Identifier03596010208255
UDI-Public03596010208255
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 11/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71420566
Device Catalogue Number71420566
Device Lot Number19KM08832
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 05/20/2020
Initial Date FDA Received06/15/2020
Supplement Dates Manufacturer Received11/04/2020
Supplement Dates FDA Received11/09/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age70 YR
Patient Weight67
-
-