• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES NAVIO SURGICAL SYSTEM AU; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES NAVIO SURGICAL SYSTEM AU; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number NPFS02070
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/19/2020
Event Type  malfunction  
Event Description
It was reported that during a preventive maintenance, the fully charged unit was able to boot and run for 1.5 hrs with no issues.Upon shutdown and reboots the error code 2000000000000 was continually displayed.It was tried troubleshooting, charged for an extended period, changing ups/usb ports, checking trip switches, and changing the power sources.However, there was no improvement.No patient was involved.
 
Manufacturer Narrative
H3, h6: the navio surgical system au, part number npfs02070, serial (b)(6) and intended to be used for treatment, was returned for evaluation.A relationship between the reported event and the device was established.The reported problem could not be visually confirmed.A functional evaluation was performed.The reported problem was confirmed.The customer ups was connected to a known-good navio system and a 2040000000000 error was outputted on the screen.A review of manufacturing and service records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints concluded this was an isolated event.The most probable cause of the reported problem is a failing ups battery cell.Although no further containment or corrective action is recommended or required at this time, all complaints are monitored and trended through post market surveillance activities.Additional information: d8/d9.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVIO SURGICAL SYSTEM AU
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
MDR Report Key10160624
MDR Text Key195381368
Report Number3010266064-2020-01599
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
PMA/PMN Number
K191223
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 11/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberNPFS02070
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/30/2020
Distributor Facility Aware Date05/19/2020
Initial Date Manufacturer Received 05/19/2020
Initial Date FDA Received06/16/2020
Supplement Dates Manufacturer Received11/01/2021
Supplement Dates FDA Received11/02/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-