Model Number 186770340S |
Device Problem
Migration or Expulsion of Device (1395)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 06/04/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
Device report from (b)(6) reports an event as follows: it was reported that on (b)(6) 2020 during a tlif (transforaminal lumbar interbody fusion) in the l4 right treating lumbar spinal canal stenosis that after the screw (186770340s) was inserted, a rod and a setscrew were deployed.The screw had come off the pedicle and the screwhead came off the rest of the screw section.The procedure was completed less than 30-minute surgical delay.Concomitant devices: unknown set screw (part# unknown, lot# unknown, quantity 1).Unknown rods (part# unknown, lot# unknown, quantity 1).This report is for one (1) viper prime cfx xtab 7x40mm ti.This is report 1 of 1 for (b)(4).
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Search Alerts/Recalls
|
|