SMITH & NEPHEW, INC. STEM IMPACTOR ROD; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
|
Back to Search Results |
|
Catalog Number 71364011 |
Device Problem
Material Deformation (2976)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 05/20/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that during inspection after procedure, it was noticed that the threads stripped.
|
|
Manufacturer Narrative
|
The device, used in treatment, was returned for evaluation.A visual inspection confirms the device¿s threads appear to be stripped and excessively worn.This device also has significant signs of wear/usage.The manufacturing date for this device is 2009.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
|
|