Model Number ACCORD TENSIONER
Device Problem Activation, Positioning or Separation Problem (2906)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/28/2020
Event Type  malfunction  
Event Description
It was reported that during the procedure the cable tensioner did not function properly with the locking mechanism of the device.No harm to patient or staff.Procedure concluded with smith & nephew back up device.
 
Manufacturer Narrative
The associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device does not confirm the stated failure mode.A functional evaluation of the returned device confirmed the stated failure mode.The device locking mechanism is stuck and not functioning.The device does not lock & unlock, which renders the device inoperative.The device shows signs of extensive use.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCORD TENSIONER
Type of Device
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10168539
MDR Text Key195584282
Report Number1020279-2020-02518
Device Sequence Number1
Product Code JDG
UDI-Device Identifier03596010485359
UDI-Public03596010485359
Combination Product (y/n)N
PMA/PMN Number
K123598
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 04/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberACCORD TENSIONER
Device Catalogue Number71360020
Device Lot Number17KM05765
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/17/2020
Initial Date Manufacturer Received 05/28/2020
Initial Date FDA Received06/18/2020
Supplement Dates Manufacturer Received04/13/2021
Supplement Dates FDA Received04/30/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-