This follow-up report is being submitted to relay additional information.Evaluation of the returned product/photographs provided confirmed there is debris inside the sterile packaging which is consistent with the appearance of the porous coating inside the sterile barrier and the inner sterile pouch was torn.The reported event is confirmed.Device history record was reviewed and no discrepancies were found.These products were likely conforming when they left zimmer biomet control.The likely condition of the product when it left zimmer biomet was conforming to specification.The root cause of the reported event is likely to be due to transit damage causing the porous coating to shed from the implant.This device falls within the scope of a corrective action which is to assess all current sterile barrier systems used to package products at zimmer biomet bridgend.As part of this corrective action, the pouch is being improved to use a stronger material (nylon), and foam end caps are being added.Also, the orientation the devices are packed in the shipper box is moving from vertical to horizontal, and the thickness of the shipper box has been increased.Upon investigation, it has been determined that the debris in the sterile packaging meets the acceptable criteria and product is conforming to specifications.Event is no longer considered reportable, and initial report should be voided if any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|