• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TORAX MEDICAL, INC. 1.5T LINX, 14B; ANTI-REFLUX IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORAX MEDICAL, INC. 1.5T LINX, 14B; ANTI-REFLUX IMPLANT Back to Search Results
Model Number LXMC14
Device Problem Patient Device Interaction Problem (4001)
Patient Problem No Code Available (3191)
Event Date 06/01/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4).Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The lot was not provided; therefore, the manufacturing record evaluation could not be performed.Additional information was requested, and the following was obtained: on what date did the implant take place? n/a.What is the product code for this complaint? lxmc14.Lot #? n/a.Does the patient have any of the allergies to metals? no.Is the patient currently taking currently taking steroids / immunization drugs? n/a.Did the patient have any pre-existing dysphagia or other conditions (other than gerd)? no.Was there any hiatal or crural repair done at the same time as the implant? unknown.Assuming yes.Was mesh used at time of implant? n/a.What was the reason for removal of the linx device? dr.Eggl said patient was unhappy and wanted linx removed.At the time of removal, was the device found in the correct position/geometry at the time of removal? yes.Have the symptoms resolved since the device was explanted? unknown.If information is obtained that was not available for the initial report, a follow-up report will be field as appropriate.
 
Event Description
It was reported that the linx device was removed from the patient on 6.1.20.The device was removed at the patients request because he was having issues.There is no additional information.
 
Manufacturer Narrative
(b)(4).Date sent: 07/29/2020.Device analysis: overall review of the device function and dimensions show no anomalies from a device that has been reasonably changed as part of the explant procedure.Visual analysis was consistent with an explanted device.The unit was returned with bent links.Link length and tensile force were found to meet the applicable specifications.Overall, no analysis conclusions relevant to the patient experience were found.The lot/batch was not provided; therefore, the manufacturing records could not be reviewed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1.5T LINX, 14B
Type of Device
ANTI-REFLUX IMPLANT
Manufacturer (Section D)
TORAX MEDICAL, INC.
4188 lexington avenue north
shoreview MN
MDR Report Key10179074
MDR Text Key199696413
Report Number3008766073-2020-00081
Device Sequence Number1
Product Code LEI
UDI-Device Identifier00855106005349
UDI-Public00855106005349
Combination Product (y/n)N
PMA/PMN Number
P100049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 06/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLXMC14
Device Catalogue NumberLXMC14
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/23/2020
Initial Date Manufacturer Received 06/01/2020
Initial Date FDA Received06/22/2020
Supplement Dates Manufacturer Received07/23/2020
Supplement Dates FDA Received07/29/2020
Patient Sequence Number1
-
-