• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC TORIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC TORIC INTRAOCULAR LENS Back to Search Results
Model Number VTICMO13.7
Device Problem No Apparent Adverse Event (3189)
Patient Problem No Patient Involvement (2645)
Event Date 06/06/2019
Event Type  malfunction  
Manufacturer Narrative
This product is manufactured in the u.S.But not marketed in the u.S.Work order search: no similar complaint was reported for units within the same lot.Claim # (b)(4).
 
Event Description
The reporter indicated the surgeon was loading a 13.7mm vticmo13.7 implantable collamer lens, -12.00/+2.5/098 (sphere/cylinder/axis), in the cartridge and the lens tore.There was no patient contact.Another same model/length lens was implanted on (b)(6) 2019 and the problem was resolved.Cause of the event was unknown.
 
Manufacturer Narrative
Additional information: h3- device evaluation: lens was returned in a micro centrifuge vial with moisture on the lens.Visual inspection found the haptic torn.Claim# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Type of Device
PHAKIC TORIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia, ca
MDR Report Key10189346
MDR Text Key198141600
Report Number2023826-2020-01280
Device Sequence Number1
Product Code QCB
Combination Product (y/n)N
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Type of Report Initial,Followup
Report Date 06/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2022
Device Model NumberVTICMO13.7
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/27/2020
Initial Date Manufacturer Received 06/01/2020
Initial Date FDA Received06/24/2020
Supplement Dates Manufacturer Received08/11/2020
Supplement Dates FDA Received09/10/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
CARTRIDGE MODEL SFC-45 - LOT # UNK; FOAM TIP PLUNGER MODEL FTP - LOT # UNK; INJECTOR MODEL MSI-PF - LOT # UNK
Patient Age23 YR
-
-